COMOSAR E-Field Probe Calibration Report

SAR_2

This document presents the calibration report for the MVG COMOSAR Dosimetric E-Field Probe, serial number 4024-EPGO-442. The calibration was performed by MVG on October 4, 2024, using the CALIPROBE test bench. The report details the methods and equipment used, adhering to IEC/IEEE 62209-1528 and FCC KDB865664 D01 standards.

The probe's sensitivity, linearity, and isotropy were evaluated, and the calibration results are presented for both air and liquid. The measurement uncertainty is estimated to be +/-11% for the frequency range 150-450MHz and +/-14% for the frequency range 600-7500MHz. Verification results are also included, demonstrating the probe's performance within specified limits.

The report includes a list of equipment used during the calibration process, ensuring traceability to the International System of Units (SI). This calibration is essential for accurate SAR measurements using the MVG COMOSAR system.


File Info : application/pdf, 64 Pages, 5.51MB

Download
Microsoft Office Word 2007
MVG: COMOSAR, Dosimetric E-Field Probe, SSE2, 4024-EPGO-442, ACR.278.12.24.BES.A, 4298.01