This document presents the calibration certificate for the ES3DV3 probe, serial number 3295, as calibrated by the MORLAB Communications Technology Co., Ltd. The calibration was performed on July 17, 2024, following the FF-Z11-004-02 calibration procedures for dosimetric E-field probes. The certificate details the traceability to national standards and provides measurement uncertainties with confidence probability.
The calibration process involves assessing various parameters, including sensitivity in free space (NORMx,y,z), sensitivity in tissue simulating liquid (ConvF), diode compression point (DCP), and modulation dependent linearization parameters. The probe's performance is evaluated in head tissue simulating media at different frequencies, and the conversion factors (ConvF) are determined. The document also includes frequency response charts, receiving patterns, and dynamic range assessments.
The calibration adheres to IEEE Std 1528-2013, IEC 62209-1, IEC 62209-2, and KDB 865664 standards. The reported uncertainty of measurement is stated as the standard uncertainty multiplied by the coverage factor k=2, corresponding to a coverage probability of approximately 95%.
File Info : application/pdf, 10 Pages, 5.75MB