This SAR (Specific Absorption Rate) test report details the evaluation of the UniRC 7 Pro Ground Control Station, manufactured by Siyi Technology (Shenzhen) Co., Ltd., for RF exposure compliance. The device was tested by Shenzhen LGT Test Service Co., Ltd. in May 2025, and the report was issued on May 22, 2025. The highest SAR value recorded was 0.989 W/kg for body exposure.
The report includes a comprehensive overview of the testing methodology, equipment used, and results obtained. It covers aspects such as EUT (Equipment Under Test) description, test environment, SAR measurement system, tissue simulating liquids, system validation, SAR evaluation procedures, antenna location, test positions, uncertainty analysis, conducted power measurements, and SAR result summary. The device supports 2.4G/5.8G/WLAN/BT modes, and the report details the SAR testing for each mode.
The report also includes appendices with system validation plots and SAR test plots, providing detailed graphical representations of the SAR distribution within the phantom. The test results demonstrate that the UniRC 7 Pro Ground Control Station complies with the specified SAR limits for general population/uncontrolled exposure.