This document presents a collection of SAR (Specific Absorption Rate) reference dipole calibration reports for MVG COMOSAR devices. These reports detail the calibration process and results for dipoles operating at various frequencies, including 835 MHz, 1800 MHz, 1900 MHz, and 2450 MHz. The calibrations were performed by MVG using the COMOSAR test bench, ensuring traceability to national metrology institutions.
The reports adhere to the requirements set forth by the IEC/IEEE 62209-1528 and FCC KDB865664 D01 standards for reference dipoles used in SAR measurement system validations. Key aspects covered include mechanical dimensions, S11 parameter (return loss), and SAR measurements in head liquid. The results demonstrate that the dipoles meet the specified criteria for return loss, impedance, and SAR values, ensuring their suitability for accurate SAR measurements.
Each report includes a list of equipment used during the calibration process, along with their respective calibration dates. The calibration results are essential for maintaining the accuracy and reliability of SAR measurement systems used in evaluating the RF exposure characteristics of wireless devices.