MVG COMOSAR SAR Reference Dipole Calibration Report

SAR Test report 3

This document summarizes the calibration results for MVG COMOSAR reference dipoles, adhering to IEC/IEEE 62209-1528 and FCC KDB865664 D01 standards. These dipoles are used for SAR measurement system validations, and the report verifies their compliance with established standards.

The calibration process involves mechanical dimension checks, S11 parameter measurements, and SAR evaluations in head liquid. The mechanical dimensions are verified using ISO17025 calibrated calipers. S11 parameters are measured against a liquid-filled flat phantom using a network analyzer. SAR measurements are performed to ensure the dipole produces the expected SAR values within specified uncertainty limits.

The report includes detailed results for each dipole, including mechanical dimensions, S11 parameter plots, and SAR values normalized to 1W forward power. Measurement uncertainties are also provided, ensuring the reliability of the calibration results. The calibration is performed by MVG, an accredited laboratory, and all results are traceable to national metrology institutions.

The tested models include SID1800, SID1900 and SID2450. The serial numbers are SN 03/15 DIP1G800-349, SN 03/15 DIP1G900-350 and SN 03/15 DIP2G450-352 respectively.


File Info : application/pdf, 25 Pages, 4.99MB

Download
Aspose.Words for .NET 15.1.0.0
tags: ["COMOSAR", "SID1800", "SID1900", "SID2450", "DIP1G800-349", "DIP1G900-350", "DIP2G450-352", "IEC/IEEE 62209-1528", "FCC KDB865664 D01", "ACR.53.26.24.BES.A", "ACR.53.27.24.BES.A", "ACR.53.29.24.BES.A"]
Array
(
    [0] => Array
        (
            [url_sha] => 91ff973f3b6ebd17cd8df992a888122ae7f01b8b4ba337cdda52131653432120
            [source] => fccid
            [url] => https://apps.fcc.gov/eas/GetApplicationAttachment.html?id=8086782
            [href_text] => SAR Test Report 3
            [href_description] => Trackimo INC. TRKM010-C Tracki 4G 2AAI6-TRKM010-C 2AAI6TRKM010C trkm010 c
            [http_status] => 200
            [http_lastmod] => 0000-00-00 00:00:00
            [http_contenttype] => application/pdf
            [http_filename] => GetApplicationAttachment.html?id=8086782
            [http_size] => 5235516
            [file_sha256] => 7486a025455549076c82dfdfdc4d31c0ee71935a4b11f3ae31bd5befc1cebe32
            [domain] => document.fccid
        )

    [1] => Array
        (
            [url_sha] => 646761bbf2da212396d4d07e8fe11f430238969cc2cf11279cbfa5fa255b9185
            [source] => fccid
            [url] => https://apps.fcc.gov/eas/GetApplicationAttachment.html?id=8086522
            [href_text] => SAR Test report 3
            [href_description] => Trackimo INC. TRKM010-C Tracki 4G 2AAI6-TRKM010-C 2AAI6TRKM010C trkm010 c
            [http_status] => 200
            [http_lastmod] => 0000-00-00 00:00:00
            [http_contenttype] => application/pdf
            [http_filename] => GetApplicationAttachment.html?id=8086522
            [http_size] => 5235516
            [file_sha256] => 7486a025455549076c82dfdfdc4d31c0ee71935a4b11f3ae31bd5befc1cebe32
            [domain] => document.fccid
        )

)
Array
(
    [file_sha256] => 7486a025455549076c82dfdfdc4d31c0ee71935a4b11f3ae31bd5befc1cebe32
    [cleanFccId] => 2AAI6TRKM010C
    [title] => MVG COMOSAR SAR Reference Dipole Calibration Report
    [description] => Accredited SAR reference dipole calibration report for COMOSAR models. Compliant with IEC/IEEE 62209-1528 and FCC KDB865664 D01 standards.
    [brand_name] => MVG
    [keywords] => ["COMOSAR", "SID1800", "SID1900", "SID2450", "DIP1G800-349", "DIP1G900-350", "DIP2G450-352", "IEC/IEEE 62209-1528", "FCC KDB865664 D01", "ACR.53.26.24.BES.A", "ACR.53.27.24.BES.A", "ACR.53.29.24.BES.A"]
    [intro] => 

This document summarizes the calibration results for MVG COMOSAR reference dipoles, adhering to IEC/IEEE 62209-1528 and FCC KDB865664 D01 standards. These dipoles are used for SAR measurement system validations, and the report verifies their compliance with established standards.

The calibration process involves mechanical dimension checks, S11 parameter measurements, and SAR evaluations in head liquid. The mechanical dimensions are verified using ISO17025 calibrated calipers. S11 parameters are measured against a liquid-filled flat phantom using a network analyzer. SAR measurements are performed to ensure the dipole produces the expected SAR values within specified uncertainty limits.

The report includes detailed results for each dipole, including mechanical dimensions, S11 parameter plots, and SAR values normalized to 1W forward power. Measurement uncertainties are also provided, ensuring the reliability of the calibration results. The calibration is performed by MVG, an accredited laboratory, and all results are traceable to national metrology institutions.

The tested models include SID1800, SID1900 and SID2450. The serial numbers are SN 03/15 DIP1G800-349, SN 03/15 DIP1G900-350 and SN 03/15 DIP2G450-352 respectively.

)
Array
(
    [0] => Array
        (
            [id] => 8086522
            [file_sha256] => 
            [applicationId] => DH2mWVSe5hzP1nWLgDqvtQ==
            [description] => SAR Test report 3
            [shortTermConfidential] => No
            [permanentConfidential] => No
            [supercede] => No
            [exhibitType] => RF Exposure Info
            [fileType] => Adobe Acrobat PDF
            [displayType] => pdf
            [fileSize] => 5235518
            [submissionDate] => 2025-03-03 00:00:00
            [dateAvailable] => 2025-03-03 00:00:00
            [creationDate] => 
            [producer] => 
            [modDate] => 
            [title] => 
            [creator] => 
            [author] => 
            [pages] => 
            [realSize] => 
            [html] => 0
            [png] => 0
            [txt] => 0
            [version] => 1
            [source] => 0
            [fccId] => 2AAI6-TRKM010-C
            [cleanFccId] => 2AAI6TRKM010C
            [exif] => 0
        )

)