EX3DV4 Probe Calibration Parameters and Modulation Details

IHDT56AV4_Part 1_SAR_Appendix H_Reference SAR_Part8-11

This document provides a comprehensive overview of the calibration parameters for the EX3DV4 probe, specifically serial number SN:7764. It includes detailed information on the probe's performance in head tissue simulating media, covering parameters such as relative permittivity, conductivity, and ConvF values for different frequencies. The calibration parameters are essential for accurate measurements in various wireless communication systems.

Furthermore, the document outlines the modulation calibration parameters for a wide range of communication systems, including GSM, GPRS, EDGE, UMTS, LTE, and WLAN. Each system is characterized by its unique identifier (UID), communication system name, group, PAR (Peak-to-Average Ratio) in dB, and associated uncertainty. This detailed information is crucial for ensuring the probe's accurate response across diverse modulation schemes and signal types.

The document also includes assessments of frequency response, receiving patterns, dynamic range, and isotropy in liquid, providing a thorough characterization of the probe's performance characteristics. Uncertainty assessments are provided for various parameters, ensuring the reliability and accuracy of measurements performed with the EX3DV4 probe.


File Info : application/pdf, 17 Pages, 3.67MB

Download
福昕风腾PDF套件 打印机版本 7.3.4.0308
SPEAG: EX3DV4, SN:7764, EX-7764_Sep24