This document presents the method and results from an accredited SAR reference dipole calibration performed by MVG using the COMOSAR test bench. The report details calibrations for reference dipoles operating at 750 MHz, 835 MHz, and 1800 MHz. These dipoles are designed in accordance with IEC/IEEE 62209-1528 and FCC KDB865664 D01 standards and are intended for use with the COMOSAR test bench.
The calibration process ensures that the dipoles meet specific mechanical and electrical requirements, including S11 parameter and SAR performance. Measurements are traceable to national metrology institutions, providing confidence in the accuracy and reliability of the calibration results. The report includes detailed information on the measurement setup, uncertainty analysis, and calibration results for each dipole, including mechanical dimensions, S11 parameter in head liquid, and SAR values with head liquid.
This report is essential for ensuring the accuracy of SAR measurements performed using MVG COMOSAR systems.